Seminario de Enfermería: Alarmas en la Ventilación Mecánica. ¿Qué significan?

Seminario de Enfermería: Alarmas en la Ventilación Mecánica. ¿Qué significan?

Introduction to Mechanical Ventilation Alarms

Overview of the Seminar

  • The seminar is introduced by a speaker who presents Abel Reyes Pérez, a university technician specializing in respiratory therapy, focusing on mechanical ventilation alarms.
  • The importance of understanding these alarms is emphasized due to the critical condition of many patients during the pandemic and the increased use of mechanical ventilators.

Purpose of the Presentation

  • The main goal is for healthcare professionals, particularly nursing staff, to identify immediate health risks related to patient conditions and ventilator malfunctions.
  • Emphasis on seeking help when necessary to ensure proper patient care and response to alarm signals from ventilators.

Understanding Ventilator Alarms

Types of Alarms

  • Ventilators display alarms in various colors: red indicates high priority issues while yellow signifies less urgent attention required.
  • Alarms can be categorized into two types: those related to patient conditions (e.g., peak pressure alarms) and technical events concerning equipment functionality.

Role of Healthcare Professionals

  • Nurses must understand what each alarm signifies so they can alert physicians for timely interventions.
  • Modern ventilators provide detailed monitoring options that allow healthcare providers to visualize pulmonary mechanics numerically or graphically.

Advanced Monitoring Features

Technological Advancements in Ventilators

  • Newer ventilator models offer advanced features like dynamic lung graphics that visually represent patient conditions, aiding in diagnosis without extensive technical knowledge.
  • These tools are especially useful for patients with chronic obstructive pulmonary disease (COPD), providing insights into their respiratory status through visual aids.

Technical Events and Challenges

  • Technical issues such as power loss or pneumatic source failures pose significant challenges; these should be communicated promptly within medical teams.

Ventilator Management and Alarm Settings

Importance of Alarm Settings

  • The ventilator may not achieve the desired volume due to low alarm settings, necessitating adjustments by medical personnel.
  • Default alarm configurations can lead to excessive noise during shifts, causing distractions without providing critical information.

Initial Setup for Mechanical Ventilation

  • Calibration is essential when initiating mechanical ventilation; respiratory therapy technicians must ensure all ventilators are functioning correctly and free from leaks.
  • It's crucial to verify that the ventilator displays proper functionality before patient connection to avoid misdiagnosing equipment failure.

Understanding Ventilator Models

  • Common models include Hamilton ventilators, which feature a proximal flow sensor located in the circuit.
  • These sensors protect the equipment from patient-related contaminants while aiding in monitoring pulmonary mechanics.

Monitoring Oxygen Levels

  • The importance of oxygen cell maintenance is highlighted; typically changed annually, these cells impact accurate oxygen delivery readings.
  • Alarms indicating low oxygen levels require immediate attention as they may not reflect actual patient oxygenation but rather sensor inaccuracies.

Troubleshooting Ventilator Issues

  • A "low volume" alarm indicates potential malfunction; if no data is displayed on the monitor, disconnecting the patient and notifying respiratory therapy is necessary.

Understanding Ventilator Functionality and Alarms

Importance of Sensor Verification

  • When a ventilator malfunctions, it may appear non-functional; it's crucial to check the flow sensor connection if the patient shows signs of thoracic distension.
  • The ventilator operates without feedback on mechanical parameters, similar to first-generation models, which can lead to complications in patient care.

Alarm System Overview

  • Ventilators utilize a color-coded alarm system: red indicates high-priority alarms requiring immediate attention, while yellow signifies medium priority or minor alerts that can be dismissed.
  • The inspiratory and expiratory branches of the ventilator include an antibacterial filter and water trap; improper drainage can cause high leak alarms.

Handling Water Traps and Circuit Issues

  • Care must be taken to ensure the water trap clamp is closed; an open clamp leads to high leak readings and potential alarm triggers.
  • If the ventilator circuit is closed but no patient is connected, a yellow alarm will indicate that ventilation has not started despite the presence of a patient.

Responding to Alarms

  • Red alarms signal critical issues such as standby mode being activated unintentionally; immediate action is required to resume ventilation.
  • Distractions during patient handling may prevent proper initiation of ventilation; vigilance is necessary when monitoring alarm statuses.

Oxygen Cell Monitoring

  • Oxygen sensors are vital for accurate readings; if they malfunction, they may report incorrect oxygen levels (e.g., 21% instead of programmed values).
  • Nurses should monitor patients for desaturation without apparent cause and notify medical staff if discrepancies arise with oxygen delivery systems.

Equipment Connectivity Concerns

  • Moving equipment can inadvertently loosen connections; this may result in reliance on internal battery power, leading to operational failures if not addressed promptly.

Ventilator Functionality and Alarm Systems

Understanding Ventilator Alerts

  • The ventilator's operational integrity is indicated by a battery icon that blinks if the power supply is disrupted or the cable becomes loose.
  • Ideally, the internal battery should show a green status; however, due to wear over time, not all batteries may function optimally.
  • The ventilator features an inspiratory branch and an expiratory branch, including a valve block composed of a valve, water trap, membrane for microorganism protection, and an expiratory flow sensor.

Sensor Flow Errors and Alarms

  • An "error de sensor de flujo" alarm indicates issues with the flow sensor component of the ventilator which is crucial for proper expiration.
  • The ventilators have color-coded alarms: red for high priority alerts indicating critical failures, yellow for medium priority issues, and informational alerts in other colors.

High Flow Models and Pressure Alarms

  • Newer models designed for high flow can trigger low-pressure alarms; these alarms can be disabled during high-flow operation to prevent unnecessary interruptions.
  • If there’s a leak detected in the system (indicated by specific codes), immediate medical attention is required as it signifies potential equipment failure.

Maintenance Protocols

  • When leaks are suspected due to alarm signals, technicians must assess whether to replace parts or monitor performance based on volume output versus programmed settings.
  • Disconnecting the ventilator inadvertently will trigger a battery mode alarm; ensure connections are secure to avoid false alarms.

Operational Guidelines for Specific Models

  • In emergency settings like urgent care where BG 70 ventilators are used, pressing standby requires holding down a button for three seconds to activate necessary screens.
  • High-priority alarms appear in red while informational ones show up in yellow; understanding these distinctions helps manage patient safety effectively.

Oxygen Supply Monitoring

  • An oxygen low alert does not immediately indicate equipment failure but suggests that oxygen pressure may be below recommended levels during calibration tests.

Ventilator Functionality and Pressure Management

Overview of Ventilator Pressures

  • The entry point of each service features a display indicating the pressure levels for oxygen and air supply, typically below 50 psi.
  • An example is given of a transport ventilator used in emergency settings, which operates at pressures significantly lower than recommended (e.g., 35 to 89 psi).

Importance of Pressure Maintenance

  • Low pressure can lead to ventilator malfunction; historical context provided regarding high contagion periods affecting functionality.
  • Biomedical maintenance checks are crucial; all institute ventilators function effectively under 50 psi.

Operational Checks for Ventilators

  • To ensure proper operation, run the ventilator for an hour with a test balloon to check for low-pressure alarms or discrepancies in oxygen fraction.
  • Monitoring equipment should confirm that the set parameters align with actual readings over time.

Understanding Alarm Systems

  • Each ventilator has specific alarms; attention must be paid to leak detection indicated by graphical information on volume curves.
  • Initial checks should include ensuring all circuits are properly connected and sealed to prevent leaks.

Managing Leak Alarms

  • If a leak alarm persists, verify that aspiration circuits are correctly sealed and maintain appropriate water column pressure (20 to 30 cm).
  • Regular monitoring is necessary as water column pressure decreases over time, impacting performance.

Troubleshooting Persistent Issues

  • If issues continue after initial checks, consider potential disconnections or filter blockages that may require medical evaluation.
  • A successful leak test indicates no internal failures; further patient assessment is needed if alarms persist despite equipment passing tests.

Apnea Alarm Considerations

Understanding Apnea Alarms in Ventilators

Overview of Apnea and Ventilator Safety Features

  • The discussion begins with the issue of apnea, emphasizing that patients can remain in this state for extended periods. Ventilators are equipped with a backup mode to activate if apnea is detected, ensuring patient safety.
  • When an apnea alarm is triggered, it serves as a critical alert for healthcare providers. It’s essential to check the ventilator settings, particularly when using pressure support modes.

Alarm Settings and Recommendations

  • The ventilator provides an automatic menu for configuring apnea alarms based on volume or pressure. Users can set specific parameters to ensure timely intervention.
  • Recommended settings suggest maintaining the apnea alarm between 15 to 20 seconds. This timeframe allows for effective backup activation without compromising patient care.

Circuit Issues and Alarm Responses

  • High peak pressures may indicate blockages due to secretions or improper intubation techniques. It's crucial to monitor these changes closely.
  • Condensation in aspiration circuits can lead to increased resistance; audible alerts (like buzzing sounds) signal potential issues that need immediate attention.

Immediate Actions During Alarms

  • In case of alarming conditions, healthcare professionals must act quickly by disconnecting the patient from the ventilator if necessary and seeking assistance immediately.
  • Some ventilators will display warnings about inadequate ventilation or malfunctioning flow sensors, which could lead to ineffective treatment if not addressed promptly.

Importance of Vigilance with Alarms

  • Continuous alarms should not be normalized; staff must remain attentive rather than desensitized to alarm sounds during shifts.
  • Familiarity with different ventilator models is vital for effective troubleshooting and reporting any malfunctions accurately to biomedical teams or respiratory therapists.

Best Practices for Alarm Management

  • Healthcare providers should avoid disabling alarms even during high-frequency alerts; doing so could mask critical changes in patient status that require intervention.
  • Always consult experienced colleagues or medical personnel when uncertain about equipment functionality—this collaboration can prevent adverse outcomes for patients.

Conclusion and Acknowledgments

Understanding Humidification and Ventilation Challenges

Issues with Humidifiers in Ventilation

  • The Hamilton ventilator may show measurement errors when humidity is present, affecting respiratory therapy. This issue arises particularly with humidifiers of the same brand.
  • A common problem is that these humidifiers can significantly increase temperature. Switching from invasive to non-invasive modes can help maintain a target temperature between 31 to 33 degrees Celsius.
  • High temperatures reported by the ventilator often relate to the humidification system's inefficiencies, which can lead to excessive heating.

Monitoring Balloon Inflation

  • Besides using a manometer, balloon inflation can be checked with a baumanometer and a three-way valve syringe. Ideal pressure should be maintained between 18 and 22 mmHg for optimal performance.

Condensation Issues in Circuits

  • Condensation in circuits can occur even without humidification due to routine saline installations or unnecessary suctioning of patients, leading to water accumulation.
  • Artificial noses function by condensing exhaled moisture; however, excess water in aspiration circuits indicates improper management practices.
  • It’s recommended to only aspirate when necessary as routine suctioning without patient assessment leads to circuit issues.

Circuit Disconnection and Patient Breathing Patterns

  • If disconnection is indicated without actual disconnection, it could signal large leaks or spontaneous breaths from the patient. Verifying patient breathing against set frequencies helps diagnose this issue.
  • To determine if a ventilator is auto-cycling due to spontaneous breaths, observe the pressure graph during expiratory pauses; negative deflections indicate patient breaths while positive ones suggest ventilator cycling.

Circuit Collapse Concerns

  • Circuit collapse may happen infrequently but could result from elevated peak pressures or obstructions within the tubing.
  • Checking for blockages or kinks in endotracheal tubes is crucial as they might cause abnormal sounds or movements in circuits under high pressure conditions.

Teamwork in Respiratory Therapy

Importance of Interdependent Activities

  • The discussion emphasizes the significance of teamwork in healthcare, particularly in respiratory therapy, where interdependent activities are crucial for patient care.
  • It highlights the responsibility of all team members to ensure that patients on mechanical ventilation maintain proper equipment function.

Challenges in Specialized Care

  • Many hospitals lack specialized inhalotherapy personnel, which can impact patient outcomes.
  • The Instituto Nacional de Enfermedades Respiratorias has a training school for specialized staff, underscoring the need for expertise in managing respiratory equipment and alarms.

Acknowledgments and Future Engagement

  • Appreciation is expressed towards participants and contributors, including Maestro Abel and the INER community, for their involvement in the seminar.